Highest Accuracy PMT Spectrometer
Specifications
  • Wavelength range: 380-800nm
  • Wavelength accuracy: +/- 0.2nm
  • Wavelength repeatability: +/- 0.1nm
  • Spectrum sample interval: 5nm, 1nm, 0.1nm
  • Chromaticity accuracy: +/- 0.0003
  • Color temperature range: 1500K - 25000K
  • Color temperature accuracy: +/- 0.3%
  • Color rendering index range: 0 - 100.0
  • Photometry linearity: 0.3%
  • Luminosity range: 0.01lm - 1.999 x 10 (to the 5th) lm with sphere
  • Photometer head accuracy: CIE Class A
  • Temperature test range: 0 - 50 degrees Celsius
  • Temperature accuracy: +/- (0.5% range + 0.1 degrees C) 

PMS-50
Highest Accuracy PMT based spectrometer

The PMS-50 is the most sensitive spectrometer available.  It is designed to be used with integrating spheres to measure integrated optical signals more accurately than other systems available.  Accuracy, linearity, and repeatablility are the key benefits of the PMS-50.
Typical test report example